Título: | ROOMTEMPERATURE PHOSPHORIMETRY BASED ON DETERGENTLESS MICROEMULSIONS: METHOD DEVELOPMENT AND APPLICATION AIMING THE DETERMINATION OF CARBAZOLE AND DIBENZOTHIOPHENE IN ORGANIC LIQUID SAMPLES | ||||||||||||||||||||||||||||||||||||||||||||
Autor: |
YANETH EVELYN VASQUEZ CASTRO |
||||||||||||||||||||||||||||||||||||||||||||
Colaborador(es): |
RICARDO QUEIROZ AUCELIO - Orientador |
||||||||||||||||||||||||||||||||||||||||||||
Catalogação: | 28/AGO/2006 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=8902&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=8902&idi=2 |
||||||||||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.8902 | ||||||||||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||||||||||
A new approach for the determination of one nitrogen-
containing PAH
(carbazole, CBZ) and one sulfur-containing PAH
(dibenzothiophene, DBT) in
organic liquid samples using room-temperature
phosphorimetry was evaluated.
Detergentless microemulsions were used to allow the
determination of these
compounds in aqueous solution. A preliminary study of the
phosphorescent
properties of the analytes was made in two steps: (i)
definition of the ME
composition and (ii) effect of the concentration of the ME
on the analyte
phosphorescent signal. As a result, phosphorescent signals
with analytical
potential, in a fluorescence free environment, were
obtained for several tested
compositions. Experimental conditions were optimized using
a factorial design
aiming the maximization of the phosphorescence from CBZ
(292/436 nm) and
DBT (286/442 nm). The concentration of I- (phosphorescence
inducer heavy atom
ion), concentration of Na2SO3, pH and ME type were
evaluated. The method
allowed, in the diluted ME, values for LD and for LQ of
2,6 ng mL-1 and 8,6
ng.mL-1for CBZ and of 7,4 ng mL-1 and 24,5 ng mL-1 for
DBT. Linear dynamic
ranges extended up to 800 ng mL-1 (CBZ) and up to 1400 ng
mL-1 (DBT).
Organic liquid samples (petroleum fuel fractions and
organic extracts of
environmental samples) spiked with the analytes of
interest were used to test the
methods. Proper strategies for analyzing each type of
sample were described.
Recoveries for CBZ and for DBT were within the 90 to 110 %
range. Mutual
interferences and interferences from other substances were
evaluated.
|
|||||||||||||||||||||||||||||||||||||||||||||
|