Título: | METROLOGICAL EVALUATION OF THE INTERLABORATORIAL COMPARISON OF THE ULTRASONIC FLOW METER CALIBRATION | ||||||||||||||||||||||||||||||||||||
Autor: |
WALACE FERREIRA DE ARANTES |
||||||||||||||||||||||||||||||||||||
Colaborador(es): |
ALCIR DE FARO ORLANDO - Orientador VALTER YOSHIHIKO AIBE - Coorientador |
||||||||||||||||||||||||||||||||||||
Catalogação: | 23/MAI/2007 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=9970&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=9970&idi=2 |
||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.9970 | ||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||
This work deals with the metrological evaluation of the
same ultrasonic flow
meter by the transit time method, when installed in two
different calibrating
systems, belonging, respectively, to two accredited
laboratories, both operating
according to the manufacture´s installation technical
specification, and following
the minimum requirements, as specified in international
standards and
recommendations. The research investigated the effect in
calibration due to the
installation of a valve and a long 90° curve, immediately
upstream of the
measurement test section, with its length dimensioned
according to the minimum
requirements of the applicable standards. It was concluded
that a longer straight
sections (at least 40 diameters) should be specified, so
that the influence of the
installation of the valve and the curve could be
neglected. The development of the
present work was motivated by the growing utilization of
ultrasonic flow meters
by the oil industry, and by the urgent need to contribute
to the advance of the
technology and the applicable standardization. In this
aspect, this work contributes
to:
(i) a better understanding of the measurement technology
and of the procedure of
ultrasonic liquids flow meters;
(ii) the advance of the technology standardization, since
it includes a critical
analysis of the minimum requirements, as specified by the
standards;
(iii) improving the laboratory accreditation criteria for
flow meter calibration,
because the influence of the installation in the
calibration was experimentally
shown.
|
|||||||||||||||||||||||||||||||||||||
|