Título: | INFLUENCE OF THE DEVELOPING VELOCITY PROFILE ON FLOW MEASUREMENT USING A TWO CHANNEL CLAMP-ON TYPE ULTRASONIC FLOW METER | ||||||||||||
Autor: |
FELIPE BORGES COELHO |
||||||||||||
Colaborador(es): |
ALCIR DE FARO ORLANDO - Orientador |
||||||||||||
Catalogação: | 23/FEV/2018 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=33101&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=33101&idi=2 |
||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.33101 | ||||||||||||
Resumo: | |||||||||||||
Industrial plants need some kind of measurement, especially chemical and oil and gas industries, which uses a large amount of flow meters, mainly those that uses ultrasonic principle due to their greater reliability. The metrological performance of this type of meter is sensitive to imperfections present in the flow profile, and therefore must comply the installation requirements relative to the distance of straight sections without objects that cause any kind of disruption in the flow. However, in some cases such recommendations are not met due to the compact physical space, which does not preclude the installation of such instruments. For this reason, the evaluation of impacts on the uncertainties of measurement caused by irregularities in the flow becomes relevant. This work shows that using a two channels clamp-on flow meter, installed in various longitudinal positions after consecutive bend sections in the pipe, it is possible to, distance starting from 20 diameters, use the meter as the standard traveling meter for calibration of others flow meters arranged in-line, properly located in the field, process called in-situ calibration. The test results provided an uncertainty less than 1 percent, which is usually specified for operational meters. This work also shows that the meter factor can be used to minimize the influence of the flow rate
fluctuation on the uncertainty of measurement, and thus to estimate more realistically its value when there is no fluctuation.
|
|||||||||||||
|