Título: | A METHODOLOGY FOR METROLOGICAL EVALUATION AND DETERMINATION THE FREQUENCY OF CALIBRATION IN FISCAL AND ALLOCATION NATURAL GAS MEASUREMENT SYSTEMS | ||||||||||||||||||||||||||||||||||||
Autor: |
JOSE LUIZ DIAS |
||||||||||||||||||||||||||||||||||||
Colaborador(es): |
ALCIR DE FARO ORLANDO - Orientador |
||||||||||||||||||||||||||||||||||||
Catalogação: | 13/DEZ/2007 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=11027&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=11027&idi=2 |
||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.11027 | ||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||
In the present work, of interest of the oil industry, a
methodology was
developed for metrological evaluating and determination
the frequency of
calibration of natural gas flow rate measurement systems
with orifice plates,
using calibration data obtained every two months from
measurement instruments
installed in fiscal points of the Campos Basin. The study
evaluates the
calibration data of temperature, pressure and differential
pressure measurement
from a multivariable transmitter, contributing to a better
understanding of the
influence of those variables in the uncertainty of flow
rate as a result of the
methodological evaluating, the intervals between two
successive calibrations
was determined, so that the requirements of the Portaria
Conjunta n° 1
ANP/INMETRO, 19/06/2000, could be met in the fiscal and
allocation points.
First of all, it was shown that the flow rate uncertainty
values, as calculated by
ISO 5167-2 standard was fully covered by the requirements.
Finally, it was
shown the interval between successive calibrations could
be increased of the
value nowadays required by portaria (2 months) to 6 months
for pressure and
differential pressure, and to 14 months for temperature.
This results has an
economic and operational impact in the natural gas
industry.
|
|||||||||||||||||||||||||||||||||||||
|