Título: | METHODS FOR THE IMIDACLOPRID QUANTIFICATION IN AQUEOUS SOLUTIONS: METROLOGICAL VALIDATION AND COMPARISON BETWEEN UV-VIS SPECTROPHOTOMETRY AND HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY | ||||||||||||||||||||||||||||||||||||
Autor: |
KELLY NEOOB DE CARVALHO CASTRO |
||||||||||||||||||||||||||||||||||||
Colaborador(es): |
REINALDO CALIXTO DE CAMPOS - Orientador RICARDO QUEIROZ AUCELIO - Coorientador |
||||||||||||||||||||||||||||||||||||
Catalogação: | 21/MAI/2007 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=9960&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=9960&idi=2 |
||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.9960 | ||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||
Fit for purpose analytical procedures must be sufficiently
reliable to support
any decision taken based on the generated results. In
order to achieve that,
consolidated adequacy evaluation of the analytical
procedure must be obtained
by performing validation experiments. In this work, an
alternative and simpler
spectrophotometric method for the quantification of
imidacloprid in aqueous
samples was compared to the HPLC-UV based reference method
used in
routine. The overall validation procedure started with a
detailed description of
each validation stage, followed by the settling of the
limits for each of the
validation parameters and then, applying the following
statistical techniques to
evaluate each of the parameters: ANOVA, regression
analysis, significance
tests, control charts and uncertainty estimation. The
measurement of
uncertainties estimation, based on ISO-GUM
recommendations, was done for
both analytical procedures (the alternative one and the
reference one). After
comparing these uncertainties with the tolerance values,
the adequacy of the
alternative proposed procedure was confirmed. Finally, by
consolidating the
validation, the experimental comparison of quantification
methods was
conducted. Besides evaluating the analytical procedure for
trace-level
imidacloprid quantification in water samples, the proposed
method was also
evaluated as the analytical procedure for imidacloprid
based formulated
products. In this case, it was demonstrated that the
spectrophotometric method
did not present the requested metrological requirements
for such application,
since the estimated uncertainties of the alternative
procedure were about 50 %
of the tolerance values.
|
|||||||||||||||||||||||||||||||||||||
|