Título: | METROLOGICAL EVALUATION OF CALIBRATION METHODS OF HIGH PRESSURE PIEZOELECTRIC TRANSDUCERS APPLIED TO BALLISTIC TESTS OF SMALL ARMS AMMUNITION | ||||||||||||
Autor: |
CAIO BITTENCOURT CARDOSO FELIX |
||||||||||||
Colaborador(es): |
CARLOS ROBERTO HALL BARBOSA - Orientador KHRISSY ARACELLY REIS MEDEIROS - Coorientador |
||||||||||||
Catalogação: | 15/JUL/2024 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=67333&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=67333&idi=2 |
||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.67333 | ||||||||||||
Resumo: | |||||||||||||
The calibration procedures of piezoelectric transducers, addressed by the
normative basis of small arms ammunition testing, lack crucial details to ensure
metrological reliability in the transient pressure measurements of these small arms
munitions. In view of this fact, it is up to the laboratories, as a requirement for
accreditation according to ISO/IEC 17025:2017 standard, to prove the effectiveness
of the methods used, especially in calibrations, which motivated the study. The
methodological approach was based on bibliographic, documentary, experimental
and laboratory research, aiming at data collection according to the stages of
evaluation of calibration methods. Initially, the evaluation of calibration certificates
of piezoelectric transducers was carried out, studying, above all, the characteristics
of the linear regression adopted by the normative basis in the calibration procedures.
At this stage, the research proposed the adoption of fit uncertainty as the main
parameter for the evaluation of piezoelectric transducers, showing their feasibility
in relation to linearity error. Then, the characterization and modeling of the
transient pressure measurement chain, composed of a piezoelectric transducer and
a charge amplifier, was performed. The modeling allowed the simulation of
different calibration methodologies, resulting in the prediction of errors, and
providing the evaluation of the influence of the charge amplifier on transient
pressure measurements. Finally, two different calibration methodologies were
compared, highlighting the importance of automated calibration systems for greater
metrological reliability.
|
|||||||||||||
|