Título: | MERCURY DETERMINATION IN GASOLINE BY COLD VAPOR AND AAS WITH DIRECT REDUCTION IN MICROEMULSION | ||||||||||||||||||||||||||||||||
Autor: |
GEISAMANDA PEDRINI BRANDAO |
||||||||||||||||||||||||||||||||
Colaborador(es): |
REINALDO CALIXTO DE CAMPOS - Orientador |
||||||||||||||||||||||||||||||||
Catalogação: | 31/MAI/2004 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4945&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4945&idi=2 |
||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.4945 | ||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||
The mercury determination in gasoline by cold vapor and
AAS, with reduction in organic liquids and in solution of
three components (microemulsion)was studied. Some
methodologies were used, with cold vapor generation and
atomic absorption spectrometer, to know: direct
determination in gasoline; direct determination in gasoline
stabilized as microemulsion, with and without
preconcentration in gold; and direct determination in
gasoline stabilized as microemulsion using a trap for
gasoline vapors before the preconcentration in gold.
Analyzing the results, it was verified that the
measurements from the gasoline stabilized as microemulsion
supplies more repetitive readings than the gasoline in
nature; moreover, it was observed that is necessary to use
preconcentration in gold, due to the lower mercury levels
in gasoline; however, it was detected that the gasoline
vapors poison the gold. Amongst the studied traps
to prevent the poisoning of the gold, the trap with
solution of K2Cr2O7/ H2SO4 was chosen, in which, the vapors
proceeding from the reaction bottle had been bubbled and
the Hg0, oxidated; later, the solution was reduced with
SnCl2/ H2SO4 and vapors sent to the gold. Thus, the results
had indicated that the method for total mercury
determination in gasoline, in which the mercury is reduced
directly from the gasoline as microemulsion, using the
K2Cr2O7/H2SO4 trap before the preconcentration in gold,
with CV-AAS detection, is recommended. After multivariate
optimization analytical curves showed coefficients of
correlation as good as 0.9999 with a corresponding
characteristic mass of 2ng of mercury. The limit of
detection obtained for this method was 0.10 ng/mL (0.14
ng/g).
|
|||||||||||||||||||||||||||||||||
|