Título: | METROLOGICAL ASSESSMENT OF THERMOELECTRIC STABILITY OF TYPE K THERMOCOUPLE | ||||||||||||||||||||||||||||||||||||
Autor: |
CESAR LEOPOLDO DE SOUZA |
||||||||||||||||||||||||||||||||||||
Colaborador(es): |
MAURICIO NOGUEIRA FROTA - Orientador ALCIR DE FARO ORLANDO - Coorientador |
||||||||||||||||||||||||||||||||||||
Catalogação: | 14/JAN/2004 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4386&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4386&idi=2 |
||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.4386 | ||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||
The present masters dissertation entitled Metrological
Assessment of Thermoelectric Stability of Type K
Thermocouple, refers to the verification
of stability deviations in the electromotive force induced
by the use of type K thermocouple with 3mm diameter mineral
insulation. Eight samples from four national thermocouples
manufacturers were analyzed, after their exposure to a
995 different to 0,5 temperature field, thus very close to its
applicability threshold (1070 Celsius degree), simulating a continuous
regime condition of use.
The samples were annealed at 1000 Celsius degree to regulate the strain
state before the tests; their physical-chemical conditions
were assessed before the essays based on chemical and
metallographic analyses. The variations in electromotive
forces induced by use were assessed based on calibrations
made before and after the exposure of the samples to the
temperature field previously determined,performed against
reference temperatures obtained with the use of six fixed-
point cells (Ag, Al, Zn, Sn, In and Ga).
The dissertation also presents a brief history on
temperature gauging and the use of type K thermocouples, in
addition to showing the basis of the theory that
supports research development, its conclusions, thus
establishing the basis for recommendations of new research
in this field.
The main result of the research carried out enabled us to
prove that, for the imposed working conditions (calibration
on fixed-point cells and exposure to temperatures of 995 Celsius degree
under a continuous regime), all the samples presented
deviations in their calibration curves which varied from
0.2 to 4.0 Celsius degree, which made explicit higher intervals than
those expected by the users. The test s lifetime was
completed with a 134 hour time interval submission to the
temperature field.
The research is relevant in the industrial and scientific
context because type K thermocouples represent
approximately 70percent of the Brazilian thermocouple market, and
because it is an expectation and a recommendation of the
National Metrology Institute of Brazil - the Instituto
Nacional de Metrologia,Normalização e Qualidade Industrial
(INMETRO).
|
|||||||||||||||||||||||||||||||||||||
|