Título: | EVALUATION OF THE PERFORMANCE OF THE METHOD OF TPH DETERMINATION (TOTAL HYDROCARBON PETROLEUM) IN SAND FOR DETECTION IN INFRA RED | |||||||
Autor: |
ALEXANDRE RODRIGUES DO NASCIMENTO |
|||||||
Colaborador(es): |
ROBERTA LOURENCO ZIOLLI - Orientador JOSE TAVARES ARARUNA JUNIOR - Orientador |
|||||||
Catalogação: | 31/OUT/2003 | Língua(s): | PORTUGUESE - BRAZIL |
|||||
Tipo: | TEXT | Subtipo: | THESIS | |||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
|||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4059&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=4059&idi=2 |
|||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.4059 | |||||||
Resumo: | ||||||||
This dissertation deals with an optimization of TPH (Total
Petroleum Hydrocarbon) analysis methodology for samples of
contaminated sands, validating the infrared detection
technique through the use of Infracal TOG/TPH
equipment. Tests were validated using Marine Fuel 380 oil,
density 0.987 g cm-3 and viscosity 5313 cP at 20 Celsius Degrees. This
oil sample was kindly supplied by Petrobras Research Centre
(CENPES), and is the same oil that leaked from a pipeline in
REDUC Refinery on January 2000, contaminating several
beaches in Guanabara Bay, including Anil and Mauá.
The optimized methodology was applied on 110 sand samples
contaminated with the aforementioned oil, and TPH were
determined within the range of 10.0 to 155.0 g kg-1.
The validation results suggested that the methodology
performance was adequate for this application. Amongst the
metrological parameters obtained from this work, the
detection limit, 9.44 mg l-1, was a plus; since it was far
below to the concentration range obtained from this samples.
|
||||||||