Título: | CALIBRATION OF A THERMOGRAPHIC CAMERA FOR PRODUCTION PLANNING | ||||||||||||||||||||||||||||||||||||||||
Autor: |
HUMBERTO SILVINO ALVES DA COSTA |
||||||||||||||||||||||||||||||||||||||||
Colaborador(es): |
ALCIR DE FARO ORLANDO - Orientador PAULO ROBERTO DA FONSECA SANTOS - Coorientador |
||||||||||||||||||||||||||||||||||||||||
Catalogação: | 11/DEZ/2007 | Língua(s): | PORTUGUESE - BRAZIL |
||||||||||||||||||||||||||||||||||||||
Tipo: | TEXT | Subtipo: | THESIS | ||||||||||||||||||||||||||||||||||||||
Notas: |
[pt] Todos os dados constantes dos documentos são de inteira responsabilidade de seus autores. Os dados utilizados nas descrições dos documentos estão em conformidade com os sistemas da administração da PUC-Rio. [en] All data contained in the documents are the sole responsibility of the authors. The data used in the descriptions of the documents are in conformity with the systems of the administration of PUC-Rio. |
||||||||||||||||||||||||||||||||||||||||
Referência(s): |
[pt] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=11021&idi=1 [en] https://www.maxwell.vrac.puc-rio.br/projetosEspeciais/ETDs/consultas/conteudo.php?strSecao=resultado&nrSeq=11021&idi=2 |
||||||||||||||||||||||||||||||||||||||||
DOI: | https://doi.org/10.17771/PUCRio.acad.11021 | ||||||||||||||||||||||||||||||||||||||||
Resumo: | |||||||||||||||||||||||||||||||||||||||||
The operating temperature increase of electric energy
production
equipments is a sign of poor performance or the need of
maintenance before
critical limits be attained. As a diagnostic tool, the
interpretation of the infrared
signal, as received by a camera that registers the image
of a target, is often used
and referred as a thermographic camera. In this work, a
methodology was
developed to interpret the infrared signal from a camera,
aiming a maintenance
planning. Initially, a device was designed to calibrate
the thermographic camera at
PUC-Rio. It consists of a cylindrical brass block, placed
inside a controlled
temperature bath, having its upper surface painted black
and placed about 3 mm
above the liquid surface of the bath. Holes were drilled
radially, slightly bellow
the block upper surface, so that its temperature could be
measured by inserted
thermocouples. Next, the instrument was calibrated with a
black body at
INMETRO. The surface emissivity was calculated as a result
of the comparison
between the calibration results in PUC-Rio and INMETRO.
After calibration, the
impact of the uncertainty of several parameters in
temperature measurement was
calculated, following the procedures that are presently
adopted by the electric
energy utility company LIGHT ENERGIA S.A., so that to
optimize the
maintenance procedure of equipments.
|
|||||||||||||||||||||||||||||||||||||||||
|